Clean Biological Safety Cabinet is essential in microbiological laboratories, especially in environments requiring strict protection measures, such as pharmaceuticals, medicine, and scientific research.
Key Features:
1. Designed based on the principles of engineering, featuring a 10° inclined work surface for enhanced user experience.
2. Air insulation design ensures no cross-contamination between internal and external air circulation, with 100% exhaust and vertical laminar negative pressure.
3. Equipped with spring-up/down movable doors on both the front and back of the workbench for flexible and convenient access.
4. Special filters are included in the ventilation system to ensure that the exhausted air meets national standards.
5. A contact switch adjusts voltage to maintain an ideal wind speed in the working area at all times.
6. Operates via an LED panel for easy control and visibility.
7. The work area is made of 304 stainless steel for durability and ease of cleaning.
Note: Custom specifications for non-standard clean benches are also available upon request.
Technical Parameters
| Parameter / Model | BSC-1000IIB2 | BSC-1300IIB2 | BSC-1600IIB2 |
|-------------------|--------------|--------------|--------------|
| Cleaning Level | 100 Level (USA Federal Standard 209E) ≥0.5 µm Granules ≤3.5 per liter | Same | Same |
| Downward Wind Speed | ≥0.35 m/s | Same | Same |
| Inlet Wind Speed | ≥0.55 m/s | Same | Same |
| Noise | ≤58 dB(A) | Same | Same |
| Half Ventilation Peak Value | ≤3 µm | Same | Same |
| Power Supply / Max Power Consumption | AC220V / 50HZ / 1KW | Same | Same |
| **Biological Safety** | | | |
| Personnel Protection | Impact type sampler: CFU ≤10; Narrow slot type sampler: CFU ≤5 | | |
| Product Protection | All sampling vessels: CFU ≤5 | | |
| Cross Contamination Protection | All double dishes: CFU ≤2 | | |
| Working Area | 1000×650×580 | 1300×650×580 | 1600×650×580 |
| Overall Dimension | 1200×750×1950 | 1500×750×1950 | 1800×750×1950 |
| Air Supply Filter Specification | 955×545×50×1 | 1290×554×50×1 | 1590×554×50×1 |
| Fluorescent Lamp / UV Lamp Specification | 20W×2 / 20W×1 | 30W×2 / 30W×1 | 30W×2 / 30W×2 |
| Illumination | ≥650 LX | ≥650 LX | ≥650 LX |
| Weight (kg) | 300 | 340 | 380 |
| Number | Single | Double | Double |
This biological safety cabinet is designed for maximum safety and efficiency, making it ideal for a wide range of laboratory applications. Whether you're conducting sensitive experiments or handling hazardous materials, this unit provides a reliable and controlled environment to ensure accuracy and compliance.